Analysis of phosphate in wastewater using an autonomous microfluidics-based analyser

Publication Type  Conference Paper
Year of Publication  2009
Authors  Cleary, J.; Slater, C.; Diamond, D.
Conference Name  In: ICEST 2009 - International Conference on Environmental Science and Technology
Conference Date  28-30 April 2009
Publisher  World Academy of Science, Engineering and Technology
Conference Location  Rome, Italy
ISBN Number  2070-3724
Key Words  RP1; RP2
Abstract  

A portable sensor for the analysis of phosphate in aqueous samples has been developed. The sensor incorporates microfluidic technology, colorimetric detection, and wireless communications into a compact and rugged portable device. The detection method used is the molybdenum yellow method, in which a phosphate-containing sample is mixed with a reagent containing ammonium metavanadate and ammonium molybdate in an acidic medium. A yellow-coloured compound is generated and the absorption of this compound is measured using a light emitting diode (LED) light source and a photodiode detector. The absorption is directly proportional to the phosphate concentration in the original sample. In this paper we describe the application of this phosphate sensor to the analysis of wastewater at a municipal wastewater treatment plant in Co. Kildare, Ireland.

URL  http://doras.dcu.ie/2969/
Export  Tagged XML BibTex